HP 37718A Instrukcja Użytkownika Strona 106

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 190
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 105
96
Making Measurements
Measuring Jitter Transfer
HOW TO: View the Results
1 Choose the results FORMAT.
If is chosen, a plot of the jitter transfer results against the
ITU-T mask is displayed.
If ischosen, the results fromwhich the graphisconstructed are
displayed:Pointnumber,Frequency, Maskamplitude (dB),JitterGain
(dB), Result.
If applicable, results 13 through 55 can be viewed on pages 2 through
5.
2 If is chosen, choose the SCALE required.
provides a vertical axis range of +5 to -60 dB and is
recommended for viewing the high frequency portion of the graph.
This allows a clearer view of the difference between the actual result
and the ITU-T pass mask.
provides a vertical axis range of +3 to -3 dB and is
recommended for viewing the low frequency portion of the graph. This
allows a clearer view of the difference between the actual result and
the ITU-T pass mask.
3 If you wish to log the jitter tolerance results to a printer, See “Logging
Jitter Transfer Results” page 114.
GRAPH
TEXT
GRAPH
WIDE
NARROW
Przeglądanie stron 105
1 2 ... 101 102 103 104 105 106 107 108 109 110 111 ... 189 190

Komentarze do niniejszej Instrukcji

Brak uwag