Using Spectrum Analyzer Features
learn How to Enter Amplitude Correction Factors
Procedure
for
Creating
Amplitude-Correction
Factors
This procedure demonstrates how to create and activate amplitude-correction
data. Detailed descriptions of the amplitude-correction functions follow this
procedure.
1. Press
(jPRESET).
NOTE
A signal is not used in this procedure for demonstrating how to create amplitude-correction data
Disconnect any cable on the cable TV analyzer input.
2. Set the center frequency to 300 MHz and the span to 500 MHz by
pressing:
3.
Press [CAL), More 1 of 4 ,
More 2 of 4 , More 3 of 4, then
Amp Cor to access the amplitude-correction menus.
3-58
4.
Press
Edit
Amp Cor to access the editing menus for
amplitude-correction factors.
5. To clear any existing amplitude-correction data, press PURGE AMP COR
two times consecutively.