This bulletin supersedes TB 9-6625-1996-35, dated 13 November 1985.
*TB 9-6625-1996-35
DEPARTMENT OF THE ARMY TECHNICAL BULLETIN
CALIBRATION PROCEDURE FOR
FREQUENCY COUNTER
HEWLETT-PACKARD
MODELS 5345A, 5345A/E28 AND 5345AOPT12
Headquarters, Department of the Army, Washington, DC
25 June 2001
Approved for public release; distribution is unlimited
REPORTING OF ERRORS AND RECOMMENDED IMPROVEMENTS
You can help improve this publication. If you find any mistakes or if you know of a
way to improve the procedure, please let us know. Mail your letter or DA Form 2028
to: Commander, U. S. Army Aviation and Missile Command, ATTN: AMSAM-MMC-
MA-NP, Redstone Arsenal, AL 35898-5230. A reply will be furnished to you. You may
also send in your comments electronically to our e-mail address:
[email protected] or FAX 256-842-6546/DSN 788-6546.
SECTION I. IDENTIFICATION AND DESCRIPTION Paragraph Page
Test instrument identification............................ 1 2
Forms, records, and reports............................... 2 2
Calibration description....................................... 3 2
II. EQUIPMENT IDENTIFICATION
Equipment required........................................... 4 3
Accessories required.......................................... 5 3
III. CALIBRATION PROCESS
Preliminary instructions.................................... 6 4
Equipment setup................................................ 7 4
Time base stability............................................. 8 5
Sensitivity.......................................................... 9 6
A12 Interface I/O................................................ 10 8
Power supply ..................................................... 11 9
Final procedure.................................................. 12 12
Przeglądanie stron 0
1 2 3 4 5 6 ... 13 14

Podsumowanie treści

Strona 1 - *TB 9-6625-1996-35

__________ This bulletin supersedes TB 9-6625-1996-35, dated 13 November 1985. *TB 9-6625-1996-35 DEPARTMENT OF THE ARMY TECHNICAL BULLETIN CALIBRATI

Strona 2

TB 9-6625-1996-35 10 a. Performance Check (1) Mount circuit board A6 (fig. 1) on extender boards. (2) Connect multimeter between A6TP6 and +

Strona 3

TB 9-6625-1996-35 11 R18+15VRETTP2 -15V RETA7 CIRCUIT BOARD S/N 1428AR24TP7 TP3R29-5.2 VRETTP6+5VRETR19A6 CIRCUIT BOARD R16R17TP1+15V RET A7 CIRCUIT

Strona 4

TB 9-6625-1996-35 12 12. Final Procedure a. Deenergize and disconnect all equipment. b. Annotate and affix DA label/form in accordance with TB 7

Strona 5

TB 9-6625-1996-35 PIN: 046669-000 By Order of the Secretary of the Army: Distribution: To be distributed in accordance with IDN 342172, requi

Strona 7

TB 9-6625-1996-35 2 SECTION I IDENTIFICATION AND DESCRIPTION 1. Test Instrument Identification. This bulletin provides instructions for the calibrat

Strona 8

TB 9-6625-1996-35 3 SECTION II EQUIPMENT REQUIREMENTS 4. Equipment Required. Table 2 identifies the specific equipment to be used in this calibratio

Strona 9

TB 9-6625-1996-35 4 SECTION III CALIBRATION PROCESS 6. Preliminary Instructions a. The instruction outlined in paragraphs 6 and 7 are preparatory

Strona 10

TB 9-6625-1996-35 5 (4) SAMPLE RATE control fully ccw. (5) CHANNEL A and B LEVEL controls to PRESET. (6) CHANNEL A and B SLOPE switches to +

Strona 11 - TB 9-6625-1996-35

TB 9-6625-1996-35 6 9. Sensitivity a. Performance Check (1) Set function generator as listed in (a) through (c) below: (a) Output Type to Un

Strona 12 - 12. Final Procedure

TB 9-6625-1996-35 b. Adjustments (1) Substitute signal generator for function generator. (2) Connect signal generator RF OUTPUT connector to

Strona 13 - Distribution:

TB 9-6625-1996-35 8 (10) Repeat (8) and (9) above as required to compensate for interaction. (11) Disconnect oscilloscope and signal generator fro

Strona 14

TB 9-6625-1996-35 9 A:00I1 CHAN A TRIG LEVEL 1.9995 2.0005 b2 B000I1 CHAN B TRIG LEVEL -2.0005 -1.9995 b3 B:00I1 CHAN B TRIG LEVEL 1.9995 2