To capture a write of known bad data to a particular
variable
The trigger specification ANDs the bad data on the data bus, write
transaction on the status bus, and address of the variable on the address bus.
1
Select the state analyzer Trigger menu.
2 Define the terms BAD_DATA, WRITE, and VAR_ADDR to represent
the bad data value, write status, and the address of the variable.
3 Under State Sequence Levels, enter the following sequence
specification:
While storing “anystate” TRIGGER on “BAD_DATA WRITE
VAR_ADDR” one time (you use the Combination trigger term to do this)
Store “anystate”
Capturing a Bad Write to a Variable
Triggering
To capture a write of known bad data to a particular variable
1–10