AutoSaturation Correction (Mariner Data Only)
Data Explorer
Software Users Guide 5-63
5
Effect on Mariner
.RST files
The AutoSaturation Correction feature is not applied to .RST
files saved from the Mariner Instrument Control Panel, even
when Saturation Correction is turned on. Saturation requires
information about the pulser frequency used to acquire the
data, and this information is not stored in .RST files saved
from the Instrument Control Panel.
Trace label When AutoSaturation Correction is turned on, spectra in the
data file are displayed with an ASC (AutoSaturation
Correction) trace label.