HP 132A Instrukcja Użytkownika Strona 187

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 330
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 186
Chapter 4 Command Reference
Introduction
Programming Guide 4-21
Stats
:CALCulate3:AVERage:COUNt:CURRent?
Queries the current count (that is, the number of data values collected for
statistical computation).
Numeric data transferred as ASCII bytes in <NR1> format.
Range is 0 to 1,000,000.
If :CALC3:AVER[:STATe] is OFF, error -221 is generated.
If the current measurement is Totalize or Voltage Peaks, 0 is returned and
error -221 is generated.
Query only.
No statistics results exist until the :CALC3:AVER:COUN:CURR? is equal to
the specified :CALC3:AVER:COUN.
Stats
:CALCulate3:AVERage[:STATe] <Boolean>
Sets or queries the statistics post-processing enable.
Single ASCII-encoded byte, 0 or 1.
A value of 0 indicates OFF; a value of 1 indicates ON.
*RST: OFF
When this enable is ON, and :TRIG:COUN:AUTO is ON, and
[:SENS]:FUNC[:ON] is not Totalize or Voltage Peak function,
then :INIT[:IMM] initiates a complete block of measurements.
See :TRIG:COUN:AUTO in this chapter for specifics.
When this enable is OFF, :INIT[:IMM] always initiates a single measurement.
Related
Front-Panel
keys
Query Response
Comments
Related
Front-Panel
keys
Query Response
Comments
Przeglądanie stron 186
1 2 ... 182 183 184 185 186 187 188 189 190 191 192 ... 329 330

Komentarze do niniejszej Instrukcji

Brak uwag