Chapter 4 Command Reference
Measurement Instructions (:CONFigure, :FETCh, :MEASure, :READ)
4-72 Programming Guide
Descriptions of the Measurement Functions<function> (Cont.)
:MEASure[:SCALar][:VOLTage]:RISE:TIME? (Cont.)
<lower_reference>
percent range: 0 to 100 [PCT]
percent resolution: 10%
voltage range: For volts if X1 Attenuation: -5.125V to +5.125V
For volts if X10 Attenuation: -51.25V to +51.25V
voltage resolution: For volts if X1 Attenuation: .005V
For volts if X10 Attenuation: .05V
default: 10 PCT
<upper_reference>
percent range: 0 to 100 [PCT]
percent resolution: 10%
voltage range: For volts if X1 Attenuation: -5.125V to +5.125V
For volts if X10 Attenuation: -51.25V to +51.25V
voltage resolution: For volts if X1 Attenuation: .005V
For volts if X10 Attenuation: .05V
default: 90 PCT
<source_list>
range: (@1)
default: (@1)
Przeglądanie stron 237