Chapter 4 Command Reference
Introduction
Programming Guide 4-61
Descriptions of the Measurement Functions<function> (Cont.)
:MEASure[:SCALar][:VOLTage]:DCYCle? [<reference>][,(@1)]
Measures Duty Cycle.
The measurement arming is coupled to auto.
The <reference> specifies the point on the pulse where the duty cycle is
determined. This point can be specified as either a percentage or an absolute
voltage. The default units are percent.
The <reference> parameter is used to configure the trigger settings. When the
<reference> is specified as a percentage (or defaulted),
auto-trigger is enabled and auto-trigger levels are set to the specified (or defaulted)
reference value. When the <reference> is specified in volts, auto-trigger is disabled
and the absolute trigger levels are set to the specified reference value.
<reference>
percent range: 0 to 100 [PCT]
percent resolution: 10%
voltage range: For volts if X1 Attenuation: -5.125V to +5.125V
For volts if X10 Attenuation: -51.25V to +51.25V
voltage resolution: For volts if X1 Attenuation: .005V
For volts if X10 Attenuation: .05V
default: 50 PCT
<source_list>
range: (@1)
default: (@1)